Table 3-5 Leakage current test results

3.5.2.2 Condition 2: Power off, reversed outlet, X-ray off

1.
2.
3.
4.
10
EVISION
Insert the black coil cord with clamp into the CHASSIS connection located on the top of the
test meter
(Figure
3-18).
Connect the Dale 601 test meter to a wall outlet.
Important! Connect the System AC power cord to the receptacle on the test meter.
Using the clamp, connect to all conductive surfaces listed in
Readings should be close to 0.0. Address high readings by checking for damaged ground wire
or internal system wiring that could be causing current to flow to the tested component or
chassis.
Condition 1:
Condition 2:
Power off,
normal outlet,
reversed outlet,
X-ray off
On the test meter, set the Outlet switch to Reversed.
Verify the system power is OFF, main circuit breaker ON.
Using the clamp, connect to all conductive surfaces.
Record the readings in
Table
Chapter 3 Installation
Condition 3:
Power off,
Power on,
normal outlet,
X-ray off
X-ray off
Leakage current
3-5.
O
XR200
/XR220
PTIMA
AMX
I
NSTALLATION
Table
3-5. Record the readings.
Condition 4:
Condition 5:
Power on,
Power on,
reversed outlet,
normal outlet,
X-ray off
X-ray on
AMX
M
ANUAL