The system must be reset using the power button each time the DAP meter is disconnected and
then reconnected with system power on. The DAP meter does not support "Hot Swapping". Failure
to reset the system in this occurrence can result in an inaccurate dose value reported by the DAP
meter.
16. Perform DAP calibration. See the System Manual, Calibration Chapter, Hardware Calibration/

4.10.4 Installing the tablet holder assembly

When the tablet holder is on hand, contact a GE service representative to install the tablet holder
assembly.
1.
2.
3.
4.
10
EVISION
DAP Calibration section, for calibration procedure.
Shut down the system and turn off the main circuit breaker.
Unpack the tablet holder assembly.
Remove the 4 screws on the lead apron hanger, and then remove the lead apron hanger from
the column.
Hang on the whole tablet holder assembly (item 2) on the column. Put the lead apron hanger
(item 3) on the tablet holder assembly in order to install the lead apron hanger and the tablet
Chapter 4 Configuration