Recall factory trim—sensor trim
The recall factory trim-sensor trim command allows the restoration of the as-shipped
factory settings of the sensor trim. You can use this command to recover from an
inadvertent zero trim of an absolute pressure unit or inaccurate pressure source.
Recall factory trim with a Field Communicator
Procedure
From the HOME screen, enter the Fast Key sequence and follow the steps within the Field
Communicator to complete the sensor trim.
Device Dashboard Fast Keys
Recall factory trim with AMS Device Manager
Procedure
1. Right click on the device and, under the Method dropdown menu, move cursor over
2. Set the control loop to manual.
3. Select Next.
4. Select Sensor Trim under Trim to recall and click Next.
5. Follow the screen prompts to recall Sensor Trim.
Recall factory trim with local operator interface (LOI)
Procedure
Refer to
Figure 5-4: Recall Factory Trim with LOI
Rosemount 3051P
Calibrate and select Restore Factory Calibration.
Figure 5-4
to recall factory sensor trim.