Calomat 7; Connecting Gas - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

8.1.4.3

CALOMAT 7

CALOMAT 7: Arrangement of the gas connections in the rack-mounted device
Figure 8-3
Design of the gas connections in CALOMAT 7
Gas connection
Sample gas
8.1.5

Connecting gas

Procedure
1. Connect the gas lines by means of suitable clamping ring screwed connections to the
associated gas inlets and outlets.
When tightening the union nuts at the screwed connections make sure that you counter
properly.
2. Check the clamping ring screwed connections for a tight fit and the pipe connections for
leaks.
Rack-mounted device
Operating Instructions, A5E31874006-12
Slot of analyzer module 1: CALOMAT 7
1
Sample gas inlet
2
Sample gas outlet
3
Not assigned
4
Not assigned
Slot of analyzer module 2
CALOMAT 7 gas connections in the rack-mounted device, rear
Material
Stainless steel material no. 1.4571
Connecting
8.1 Gas connections
External nozzle diameter
6 mm
73