Technical specifications
13.2 Technical specifications of rack-mounted housing
Electrical characteristics
Transient protection
Power unit cable
Electrical inputs and outputs
Digital outputs (DO)
Digital inputs (DI)
Service interface (rear)
Service interface (front)
Table 13-2
Rack-mounted housing: Option modules
Option modules
Option module 1.1
Option module 2.1
Option module 2.2
Table 13-3
Rack-mounted housing: Climatic conditions
Climatic conditions
Permissible operating altitude
Permissible ambient temperature
and humidity
118
It has to be ensured that the transient protection is set to a value that
does not exceed 140% of the measured peak voltage value at the
device power supply connectors.
Cable diameter: max. Ø 10 mm (3/0 AWG)
Core cross-section: max. 3 x 2.1 mm
8, with changeover contacts, parameters can be freely assigned, e.g. for measuring range
identification; max. load: 24 V AC/DC/1.7 A (total load for all 8 relay outputs in continuous
operation max. 160 W), isolated, non-sparking
8, designed for 24 V, isolated, freely parameterizable, e.g. for autoranging
Ethernet RJ 45, 100 Mbps
Ethernet RJ 45, 100 Mbps
12 digital outputs, with changeover contacts, max. load: 24 V AC/DC / 1.7 A
(total load for all 12 relay outputs in continuous operation max. 244 W), isolated, non-
sparking
8 digital inputs, designed for 24 V, isolated, freely configurable
6 analog outputs 0/4 ... 20 mA, designed for 24 V, isolated; load: R
6 digital outputs, load rating: 24 V AC/DC/1.7 A (total load for all 6 relay outputs in
continuous operation max. 122 W), isolated, non-sparking
6 analog outputs 0/4 ... 20 mA, designed for 24 V, isolated; load: R
4 analog inputs 0/4 ... 20 mA, designed for 24 V, non-isolated, internal resistance
100 Ω
4 digital inputs, designed for 24 V, isolated
3 000 m above sea level sea level
Technical specifications of analyzer modules (Page 119)
(14 AWG)
2
≤ 750 Ω
L
≤ 750 Ω
L
Rack-mounted device