Technical specifications
13.3 Technical specifications of analyzer modules
Influence variables
Accompanying gases
Supply voltage Analyzer module 24 V DC
(-5%/+10%)
Basic device
85 V AC to 264 V AC / 47 Hz to
63 Hz
Table 13-11
OXYMAT 7: Climatic conditions
Climatic conditions
Storage and transport
Max. permissible ambient temperature surround‐
ing the basic device during operation
Ambient humidity (relative humidity) during stor‐
age, transport or operation
Restriction for installing an ULTRAMAT 7 analyzer module (combined operation): 5 ... 45 °C (41 ... 113 °F)
1)
2)
Avoid condensation on the installed components!
Table 13-12
OXYMAT 7: Gas connections
Gas connections
With hoses
With pipes
122
Zero point deviation (cross-sensitivity) in accordance with Table A.1 of EN
61207-3
< 0.1% of the current span
< 0.1% of the current span
- 30 ... +70 °C (-22 ... 158 °F)
0 ... 50 °C (0 ... 122 °F)
1)
< 90%
2)
Connecting socket for 6 mm hose
Connecting socket for 6 mm pipe
Rack-mounted device
Operating Instructions, A5E31874006-12