Installing/Removing The Sample Gas Restrictor - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Connecting
8.1 Gas connections
8.1.3

Installing/removing the sample gas restrictor

Requirements
Before you install/remove the sample gas restrictor, ensure that
• the sample gas line is disconnected from the device,
• you have read the section "Combination operation (Page 33)",
• a suitable slot screwdriver is at hand.
The difference between a sample gas restrictor and a clamping screw is:
– Sample gas restrictor: slotted screw; outlet diameter 0.5 mm
– Clamping screw: Cross-tip screw; outlet diameter 2 mm. For additional information, refer
Procedure
The installation, removal and retrofitting of clamping screws or sample gas restrictors is limited
to only a few module types. You will find more detailed information on the mounting position
and the mounting conditions in the section "Structure of the gas connections (Page 34)".
1. Install the sample gas restrictor or clamping screw:
Screw the sample gas restrictor or the clamping screw into the outlet. The tightening torque
is 1 Nm.
2. Connect the sample gas line again.
3. Remove the sample gas restrictor or clamping screw:
Proceed in reverse order.
70
to the section "Structure of the gas connections (Page 34)".