2.3 Performing Measurements

Before performing measurements, be sure to read "Usage
Notes" (p. 5), and "2.1 Inspection Before Use" (p. 33),
and "2.2 Preparation for Measurements" (p. 35) and
perform the described steps.
Procedures
Pull the sensor opening
1
lever, so that the sensor head
opens.
2
Clamp the conductor being
measured.
• Clamp so that the
conductor is in the
center of the sensor
aperture (p. 26).
• Align the sensor so that
the current direction
indication corresponds
to the direction of
current flow through
the conductor to be
measured.
Performing Measurements
Opening
lever
Low-potential
side (ground
potential side)
High-
potential side
41