Troubleshooting
Symptom
Demagnetization and automatic zero-adjustment
cannot be performed.
Demagnetization or automatic zero-adjustment has not
been completed normally.
Under the following conditions, demagnetizing and
automatic zero-adjustment cannot be performed.
Alternatively, demagnetization or automatic zero-
adjustment has not been completed normally.
• The upper jaw is unlocked.
• An overload has been detected whether or not the
upper jaw is locked.
• A current exceeding 0.5 A rms has been detected
whether or not the upper jaw is locked.
 Implement the remedy described on the following
pages:
See "Demagnetizing / automatic zero-adjustment
unavailable" (p.  9 2).
Then, re-execute demagnetization and automatic
zero-adjustment.
See "Executing demagnetization and automatic zero-
adjustment" (p.  3 3).
When demagnetization or automatic zero-adjustment
has not been normally completed even with the sensor
not clamped around any conductor and the upper jaw
locked in place, the device may be malfunctioning.
 Send the device for repair.