Iwatsu SS-500 Series Instruction Manual page 53

Hide thumbs Also See for SS-500 Series:
Table of Contents

Advertisement

Use the device for measuring currents much
lower than the maximum rated current value if
an ambient temperature is relatively high or a
current to be measured can contain
frequency components other than the
fundamental.
Self-heating could cause burns or damage to the device
even if the current under measurement is lower than the
maximum rated current.
The maximum rated current is defined as a
recommended value for when a sine-wave current is
inputted to the device at the temperatures and humidity
specified for the guaranteed accuracy.
See "Frequency derating curve" (p.  7 8).
Connect the device and other probes properly
to a circuit to be measured and your
waveform measuring instrument.
Improperly connecting them could cause electric shock
or damage to the device, other probes, and instrument.
CAUTION
Measuring Currents
2
47

Advertisement