OMNI-Test Laboratories, Inc.
Manufacturer:
Model:
Serial Number:
OMNI Tracking No.:
Calibrated Orifice:
Average Gas Meter y
Factor
0.983
Calibration Date:
Calibrated by:
Calibration Frequency:
Next Calibration Due:
Instrument Range:
Standard Temp.:
Standard Press.:
Barometric Press., Pb:
Signature/Date:
Calibration Parameters
Reference Meter Pressure ("H2O), Pr
DGM Pressure ("H2O), Pd
Initial Reference Meter
Final Reference Meter
Initial DGM
Final DGM
Temp. Ref. Meter (°F), Tr
Temperature DGM (°F), Td
Time (min)
Net Volume Ref. Meter, Vr
Net Volume DGM, Vd
Gas Meter y Factor =
Gas Meter y Factor Deviation (from avg.)
Orifice dH@
Orifice dH@ Deviation (from avg.)
where:
* Reference calibration is traceable to NIST through NIST Test # 40674, Kimble ASTM E1272, or NIST traceable laboratory
** Equations come from EPA Method 5
The uncertainty of measurement is ±0.14 ft
Control No. C-SSB-0005, Effective Date: 7/7/2011
Thermal Metering System Calibration
Apex
XC-60-EP
0702003
OMNI-00371
Yes
Average
Orifice
Meter
dH@
N/A
07/16/18
B. Davis
6 months
1/16/2019
1.000
cfm
68
oF
29.92
"Hg
30.08
"Hg
7/16/2018
Run 1
540.302
5.502
5.505
0.990
0.007
N/A
N/A
1. Deviation = |Average value for all runs - current run value|
** 2. y = [Vr x (y factor (ref)) x (Pb + (Pr/ 13.6)) x (Td + 460) ]/ [Vd x (Pb + (Pd / 13.6)) x (Tr + 460)]
** 3. dH@ = 0.0317 x Pd / (Pb (Td + 460)) x [ (Tr + 460) x time) / Vr ]^2
3
/min. This is based on the reference standard having a TAR (Test Accuracy Ratio) of at least 4:1.
Y Factor
Date
y Factor
Acceptance
FALSE
Acceptable y Deviation
Maximum y Deviation
Acceptable dH@ Deviation
Maximum dH@ Deviation
Acceptance
Standard
Calibrator S/N
Run 2
Run 3
0.00
0.00
0.00
3.00
1.70
1.00
534.8
524
529.6
529.514
534.7
0
0
0
5.505
5.593
5.22
85.0
82.0
81.0
85.0
83.0
84.0
28.8
35.5
47.8
5.514
5.100
5.593
5.22
0.981
0.978
0.002
0.005
N/A
N/A
N/A
N/A
163 of 236
Previous Calibration Comparision
Acceptable
1/16/2018
Deviation (5%)
0.997
0.04985
Acceptable
Current Calibration
0.020
0.007
N/A
N/A
Acceptable
Reference Standard *
Model
Standard Test Meter
OMNI-00001
Calib. Date
30-Oct-17
Calib. Value
0.9977
Deviation
0.014
TRUE
N/A
y factor (ref)