Safety Instructions; Preparation For Use - GE Druck DPI 610 IS User Manual

Portable pressure calibrator/indicator
Hide thumbs Also See for Druck DPI 610 IS:
Table of Contents

Advertisement

HYDRAULIC CALIBRATOR VERSION

Safety Instructions

HYDRAULIC FLUID IS INJURIOUS. OBSERVE RELEVANT HEALTH AND SAFETY
PRECAUTIONS. USE APPROPRIATE PROTECTIVE BARRIERS AND EYE
PROTECTOR.
BEFORE APPLYING PRESSURE, EXAMINE ALL FITTINGS AND EQUIPMENT FOR
DAMAGE AND ENSURE THAT ALL EQUIPMENT IS TO THE CORRECT PRESSURE
RATING.
DO NOT EXCEED THE MAXIMUM WORKING PRESSURE OF THE INSTRUMENT
(INDICATED ON START-UP SCREEN AT SWITCH-ON).
Caution:
Observe absolute cleanliness when using the instrument. Severe
damage can be caused if equipment connected to this
instrument is contaminated. Connect only clean equipment to
the instrument. To avoid any contamination, an external filter is
recommended.

Preparation for Use

Fit the bleed pipe connection to the priming port using the bonded seal as
shown in Figure A1.
Make sure that the screw-press is wound fully in (clockwise).
Make sure that the isolation valve is open (fully anticlockwise).
Connect the device or system under test to the pressure port using a
bonded seal as shown in Figure A2.
MAKE SURE THAT THE CONNECTING LINE TO THE EXTERNAL DEVICE OR SYSTEM
IS CAPABLE OF WITHSTANDING THE LINE PRESSURE TO BE APPLIED.
Note: A bleed point must be provided on the external device.
K249 Issue No. 4
WARNING
WARNING
46