Advanced Task

mA Ramp

Press the OUTPUT key and select mA Ramp in a similar manner to that shown
above.
Define ramp required by entering START and END current values as shown
below (e.g.),
Connect an external power source as shown on Page 24.
Press START (F1) to run the program. A status display CHECK LOOP
indicates a fault in the external loop i.e. supply fault or open circuit .
Note: The ramp cycle (min to max or max to min), is approximately 60 seconds.
Press STOP (F1) when running to stop at any point. Press CONTINUE (F1) to
resume from point of pause or RESET (F2) to return to start point.