Taking Measurements
Leak Test (LEAK TEST) Task
External Pressure Source
(Indicator Only)
Pressure
Regulator
•
If required, use the INPUT key to change pressure units.
•
Set-up the leak test WAIT and DURATION times to the required values as
shown below. A minimum wait period of 3 minutes is recommended.
•
Close the vent valve and pressurise the external system to the required
LEAK TEST pressure.
•
Press the RUN (F2) key to start the leak test . When completed, the bleeper
sounds and the leak test results are written to the display.
K249 Issue No. 4
EXTERNAL
SYSTEM
DPI 615 IS
TASK : LEAK TEST
SNAPSHOT MODE
PRESSURE INT
WAIT
DURATION
START PRESS
STOP PRESS
PRESS CHANGE
LEAK RATE
CHANGE
VALUE
F1
16
max 30V
bar
60
secs
60
secs
bar
bar
bar
bar/m
RUN
F2