Downloading Procedure Files - GE Druck DPI 610 IS User Manual

Portable pressure calibrator/indicator
Hide thumbs Also See for Druck DPI 610 IS:
Table of Contents

Advertisement

Memory Operations
Downloading Procedure Files (DPI 615 versions only)
THE RS232 INTERFACE MUST ONLY BE USED IN A SAFE AREA
Complete test procedures may be downloaded from a PC to the DPI 615
instrument via the RS232 port. A procedure consists of a number of Druck
Universal Command Interface (DUCI) commands which are usually assembled by
a linking management software application (such as the Druck Intecal_W).
Before downloading a procedure, the DPI 615 must be in the REMOTE mode. To
place the instrument into REMOTE mode, proceed as follows.
Connect the instrument's RS232 port to a free COM port on a PC.
Make sure that the COMMS parameters of the PC match those of the
instrument (refer to Page 38).
Make sure that the instrument is not already running a procedure. If it is,
quit the procedure.
Download the procedure. Procedures are stored in the Datalog directory.
The following sequence shows a typical download sequence which commences
with the instrument in LOCAL mode.
After the procedure file has been downloaded, the instrument is usually returned
to the LOCAL mode by the last command in the procedure file. If the instrument
remains in REMOTE mode, switch it OFF and ON to reset it.
WARNING
33
Datalog
K249 Issue No. 4