Process Functions - GE Druck DPI 610 IS User Manual

Portable pressure calibrator/indicator
Hide thumbs Also See for Druck DPI 610 IS:
Table of Contents

Advertisement

Advanced Task

Process Functions

If required, the following process functions are available on the Measurand (INPUT)
display but only in ADVANCED task. If the instrument is in any other mode i.e.
BASIC or any other task mode, the input and output displays must first be
configured in ADVANCED task.
Note: PROCESS functions are not available to the output (SOURCE) channel.
A summary of the process functions follows.
Tare
Max/Min
Filter
Flow
% Span
Following selection of ADVANCED from the task menu, press the INPUT key. Use
the Up
or Down
(F1) key and use the Left
(e.g.),
Press ENTER to switch the process ON with existing settings or F1 to change
process settings (where applicable).
K249 Issue No. 4
Allows either the current display value or a manually
entered value to be tared off display parameter
reading.
Displays running Max/Min and Present display values
simultaneously. Resettable via F1 key.
Applies low pass filter function to displayed
parameter. Filter characteristics (Settling time and
Band) are user programmable.
Applies square root function to displayed parameter.
Converts displayed parameter reading to a
percentage of span. Span definable via the F1 key.
cursor keys to select the required input. Press the PROCESS
or Right
cursor keys to enable the process on/off
20
Process Functions

Advertisement

Table of Contents
loading

This manual is also suitable for:

Druck dpi 615 is

Table of Contents