Taking Measurements
Transmitter Simulator (TX SIM) Task
When used with an external voltage source (see Page 24), provides a current
output proportional to the calibrator's measured output pressure (indicated
pressure on indicator only version). Select task TX SIM. Press EXIT to skip set-up
screen if parameters are correct.
On completion of Tx SIM set-up, the display is configured as follows (e.g.).
Connect an external power source to the output loop as detailed on Page 24.
To subsequently change any of the Tx SIM scaling parameters, press CHANGE
VALUE key (F1) to obtain the TX Simulation set-up display.
To change the pressure units, press INPUT and select the required scale units. If
the required scale units are not listed, press SETUP, select SETTINGS and proceed
as detailed on Page 37.
17
K249 Issue No. 4