Running Procedure Files - GE Druck DPI 610 IS User Manual

Portable pressure calibrator/indicator
Hide thumbs Also See for Druck DPI 610 IS:
Table of Contents

Advertisement

Memory Operations
Datalog
Running Procedure Files (DPI 615 versions only)
To run a procedure, make sure that the DPI 615 is set to the Datalog mode (see
page 36), and proceed as follows.
After selecting F1, proceed by entering the User ID and Serial Number and then
select F1 (Continue) and follow the on-screen procedural instructions (e.g.),
When the test procedure for a given UUT has been completed, the result of
running the first test is stored as an AS FOUND file. This file cannot be overwriten.
Any other tests on the device ars stored as an AS LEFT file which is overwritten
each time the procedure on this device is run.
When recalling the results of a procedure, the choice of AS FOUND or AS LEFT is
provided (refer to page 35).
34
K249 Issue No. 4