Task Selection
Cal Mode (DPI 615 versions only)
Cal mode, which is available in tasks P-I, P-P, P-V, P-P, P-DISPLAY and P-SWITCH,
provides a method of setting up test parameters manually. Downloaded test
procedures can also automatically set up and turn on the Cal Mode function. The
method of turning on and setting up Cal Mode is shown below for a P-I task. The
method is similar for all the other tasks to which Cal Mode applies.
Pressing the F1 key (TURN ON CAL MODE), provides the set-up screen for the CAL
mode. Initially, the cursor is placed in the UUT SPAN field to allow the required
span range to be entered. The corresponding values for the UUT output
parameter (current) are then set, followed by the maximum error value and error
type (%rdg or %Span). When all test parameters have been set-up, the screen
changes to display the input and output and the test results. The test result can
only be displayed to within a range of ±9.99%. If the test result is outside this
range, either the left pointing (-ve error) or right pointing (+ve error) chevrons are
displayed. Within this error band, the actual tolerance value is displayed. Test
results can either be stored as snapshots or logged as datalog files, depending
upon how the instrument has been set-up.

Basic Mode (Task BASIC)

This instrument will power-up in this mode the first time that it is used. To select BASIC
from any other task, press the TASK key and select BASIC and press the ENTER key.
BASIC mode is fully described in the Getting Started, section (see page 7).