Taking Measurements

Relief Valve Test (REL VALVE) Task

To carry out a relief valve test , press TASK and select REL VALVE. Connect the
output pressure port of the instrument to an external system as shown below.
External Pressure Source
(Indicator Only)
Pressure
Regulator
To change the pressure units, if required, press INPUT and select the
required units using the cursor keys.
Close the vent valve and, using the hand-pump or external pressure
supply, apply pressure to the relief valve under test.
When the relief valve operates, the maximum recorded pressure indicates
the operating point of the valve.
Record the test results.
Note: The STORE key can be used for this purpose. Use right cursor key
initially, followed by up/down keys to enter Snapshot text.
Open vent valve to release test pressure.
Note: If using external pressure supply, isolate supply before opening the
vent valve.
K249 Issue No. 4
DPI 615 IS
TASK : RELVALVE
SNAPSHOT MODE
MAXIMUM
MINIMUM
PRESSURE INT
PRESSURE INT
RESET
MAX/MIN
F1
18
max 30V
bar
bar
F2