Teledyne 300 Instruction Manual page 18

Carbon monoxide analyzer
Hide thumbs Also See for 300:
Table of Contents

Advertisement

A critical flow orifice is used to control the sample flow. The orifice is a precision-
drilled sapphire jewel protected by a 20 micron sintered filter. The critical flow orifice
never needs adjustment and maintains precise flow control as long as the ratio of the up-
stream to down-stream pressures is greater than .53 (sonic flow conditions).
1.6.6 Sample gas connection (see Figure 1.4)
A 2 m section of 1/4" O.D. PTFE tubing is needed to connect the sample source to the
Analyzer.
USE PTFE, GLASS, STAINLESS STEEL OR
NON-REACTIVE MATERIALS FOR SAMPLE GAS
USE VENT LINE WHEN SAMPLING FROM
PRESSURIZED MANIFOLDS - SAMPLE PRESSURE
1.6.7 Zero/span valve connections (see Figure 1.4)
Zero air and span gas manifolds should supply their respective gases in excess of
800cc/min demand of the Analyzer. The manifold should be vented to the outside
atmosphere and be of sufficient length and diameter to prevent back diffusion and
pressure effects.
Adequate zero air can be supplied by the M300 using the IZS option or by using a
commercial calibrator.
P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 12
NOTE
CONNECTIONS.
NOTE
SHOULD NOT EXCEED ±1" H
O
2

Advertisement

Table of Contents
loading

Table of Contents