Signal
0
DISP_BROWNOUT
1
EXT_ZERO_CAL
2
EXT_SPAN_CAL
3
SYNC_OK
4
SPAN_VALVE
5
CAL_VALVE
6
BENCH_HTR
7
WHEEL_HTR
8
DARK_CAL
9
ELEC_TEST
10
BROWNOUT_RST
11
ST_ZERO_CAL
12
ST_SPAN_CAL
P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 52
Contro
l
NO
Display brownout is used to keep the
display from getting corrupted during
low line voltage conditions. Circuitry
on the Power Supply board (00015)
senses low line voltage and sets this
bit. The CPU reads this and generates
the
BROWNOUT_RST
described below.
NO
Shows state of status input bit to cause
the M300 to enter Zero Calibration
mode. Use to check external contact
closure circuitry.
NO
Shows state of status input bit to cause
the M300 to enter the Span Calibration
mode. Use to check external contact
closure circuitry.
NO
Indicates that demodulation circuitry
on the Sync/Demod Board (00798) is
able to lock-in on the detector signal.
YES
Switches the Zero/Span valve. Use
this bit to test the valve function.
YES
Switches the Sample/Cal valve. Use
this bit to test the valve function.
YES
Shows the status of the optical bench
heater. This has the same function as
the LED in the power supply module.
YES
Shows the status of the filter wheel
heater. This has the same function as
the LED in the power supply module.
YES
Turns off the detector input to the
Sync/Demod board for electronics
calibration.
YES
Activates the Electric Test diagnostic
circuitry.
YES
Resets
circuitry described above.
YES
Status Bit - Zero Calibration mode
Logic high = M300 in Zero cal mode
Logic low = Not in Zero cal mode
YES
Status Bit - Span Calibration mode
Logic high = M300 in Span cal mode
Logic low = Not in Span cal mode
Description
the
DISP_BROWNOUT
signal