Teledyne 300 Instruction Manual page 52

Carbon monoxide analyzer
Hide thumbs Also See for 300:
Table of Contents

Advertisement

Name
NONE
COMEAS
COREF
SLOPE1
SLOPE2
OFSET1
OFSET2
ZSCNC1
ZSCNC2
CONC1
CONC2
STABIL
BENCHTMP
WHEELTMP
SMPFLW
SMPPRS
BOXTMP
DCPS
Example:
Sample the CO concentration (Range #1) once per minute (default) and issue a report
once per 5 minutes, and print the report on the RS-232 channel.
1. Press SETUP-DAS-EDIT-ENTR-INS to create a new data channel and begin
editing its properties.
2. Edit the NAME property and give it any name you want, but do not set the blank
name (all dashes) or duplicate names because they won't be accessible from the RS-
232 interface.
3. For each of the remaining properties, press the SET> to view the current setting, and
then press EDIT to change the property if necessary. The table below lists the proper
settings for all of the properties for this example:
P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 46
Data Parameters
Description
CO Measure signal
CO Reference signal
Slope of Range #1
Slope of Range #2
Offset of Range #1
Offset of Range #2
Range1 zero/span measured
concentration
Range2 zero/span measured
concentration
Concentration of Range #1
Concentration of Range #2
Stability
Optical bench temperature
Filter wheel temperature
Sample flowrate
Sample pressure
Box temperature
DC power supply composite voltage
Units
mV
mV
mV
mV
PPM
PPM
PPM
PPM
PPM
o
C
o
C
cc/min
in-Hg
o
C
mV