CHECK THAT ANALYZER IS SET UP FOR PROPER
VOLTAGE AND FREQUENCY.
POWER PLUG MUST HAVE GROUND LUG.
7.
Turn on the M300 by switching the switch on the lower right corner of the
front panel (See Figure 2.1). The front panel display should light with a
sequence of messages, -API - M300 - software version number, then a
normal display as shown in Figure 2.2.
8.
Allow about 60 minutes for the temperatures to come up to their
respective setpoints then press the left most button on the front keyboard
to scroll through the TEST values. Compare these values to those noted
during the final factory checkout listed in Table 1.1. The values observed
should closely match the Table 1.1 values.
9.
Select the range on which the analyzer will be calibrated.
a.
From the SAMPLE menu press SETUP to enter the SETUP menu.
(See Figure 2.2 for appearance of front panel.)
b.
Enter the PASSWORD (818).
c.
Press RNGE (RANGE).
d.
Press SET.
e.
Enter the derived full scale range for analog outputs and press ENTR.
f.
Press EXIT 2 times to return to the SAMPLE menu.
10.
Adjust the analyzer zero point.
a.
Input zero air into the sample port.
b.
Press CAL from the SAMPLE menu and enter the password (818).
c.
Press ZERO.
d.
It usually takes about 5 to 10 minutes for the reading to stabilize
near zero. After a stable reading has been obtained press ENTR.
The display should now read 0.0 ppm carbon monoxide.
P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 6