Teledyne 300 Instruction Manual page 59

Carbon monoxide analyzer
Hide thumbs Also See for 300:
Table of Contents

Advertisement

13
ST_FLOW_ALARM
14
ST_TEMP_ALARM
15
ST_DIAG_MODE
16
ST_POWER_OK
17
ST_PRESS_ALARM
18
ST_HIGH_RANGE
19
ST_SYSTEM_OK
20
ST_BENCH_ALARM
21
ST_SOURCE_ALARM
22
ST_WHEEL_ALARM
23
CO_MEASURE
24
CO_REFERENCE
25
SAMPLE_PRES
26
SAMPLE_FLOW
27
SAMPLE_TEMP
P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 53
YES
Status Bit - Flow alarm
Logic High = Sample flow out of spec
Logic Low = Flows within spec
YES
Status Bit - Temperature alarm
Logic High = One or more temps out
of spec
Logic Low = Temps within spec
YES
Status Bit - In Diagnostic mode
Logic High = M100A in Diagnostic
mode
Logic Low = Not in Diag mode
YES
Status Bit - Power OK
Logic High = Instrument power is on
Logic Low = Instrument power is off
YES
Status Bit - Flow alarm
Logic High = Sample pressure out of
spec
Logic Low = pressure within spec
YES
Status Bit - Autorange High Range
Logic High = M300 in high range
Logic Low = M300 in low range
YES
Status Bit - System OK
Logic High = No instrument warning
present
Logic Low = 1 or more alarm present
YES
Status Bit - Bench Temperature Alarm
Logic High = Bench Temp out of spec
Logic Low = Bench Temp in spec
YES
Status Bit - IR Source Alarm
Logic High = IR Source output too
low
Logic Low
normal
YES
Status Bit - Wheel Temperature Alarm
Logic High = Wheel Temp out of spec
Logic Low = Wheel Temp in spec
NO
IR detector reading during measure
phase. Typically 2500-4500 mV
NO
IR detector reading during reference
phase. Typically 2500-4500 mV
NO
Sample pressure in mV. Typical sea
level value = 4300mV for 29.9" HG-
A.
NO
Sample flow in mV.
NO
Sample temp in mV