Teledyne 300 Instruction Manual page 53

Carbon monoxide analyzer
Hide thumbs Also See for 300:
Table of Contents

Advertisement

Property
NAME
EVENT
PARAMETERS
REPORT PERIOD
NUMBER OF
RECORDS
RS-232 REPORT
CHANNEL
ENABLED
CAL. HOLD OFF
*
Sample measurement period of filtered CO concentration is 1 minute by default.
4.11.3 RS-232 reporting
Automatic RS-232 reporting can be independently enabled and disabled for each Data
Channel. For all default data channels, RS-232 reporting is initially set to "OFF." If this
property is turned on, the Data Channel will issue a report with a time and date stamp to
the RS-232 port every time a data point is logged. The report format is shown below:
D 31:10:06 0412 CONC : AVG COCNC1=6.8 PPB
The report consists of the letter "D" follow by a time/date stamp("31:10:06") followed by
the instrument ID number("0412"). Next is the Data Channel name("CONC") and the
sampling mode ("AVG" indicates that the data point is an average of more than one
sample as opposed to a instantaneous reading, "INST").
parameter and its value("COCNC1=6.8 PPB") are printed. For Data Channels that
sample more than one parameter, such as PNUMTC and CALDAT, each parameter is
printed on a separate line.
To enable RS-232 reporting for a specific Data Channel:
P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 47
Setting
"CO5MIN"
ATIMER
CONC1,AVG
000:00:05 (i.e. 5 minutes)
4032 (i.e. two weeks)
ON
ON
ON
DESCRIPTION
Sample when automatic
timer is expired
SO
Range #1 concentration
2
Average of 5 minutes
reported every 5 minutes
Stores 4032 of Report
Period (5 minutes average)
RS-232 is enabled
Data channel is enabled
Hold off is enabled
Finally, the name of the
*