These commands may be used whether the diagnostics have been entered from the
keyboard (SETUP-DIAG) or the RS-232 ("D ENTER <CRLF>"). However, when the
diagnostics are entered via the keyboard, no feedback is sent to the RS-232 channel. This
prevents the RS-232 output from getting unnecessarily cluttered with diagnostic data.
Whenever the diagnostic mode is entered or exited, a report is issued to the RS-232
output. The table below summarizes the diagnostic reports.
"C DDD:HH:MM IIII ENTER DIAGNOSTIC MODE"
"C DDD:HH:MM IIII EXIT DIAGNOSTIC MODE"
P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 61
NOTE
THE DIAGNOSTICS MODE MAY ONLY BE
ENTERED VIA THE RS-232 PORT WHEN
THE ANALYZER IS IN SAMPLE MODE.
DIAGNOSTIC COMMANDS
TABLE 7.5
DIAGNOSTIC REPORTS
Report
TABLE 7.6
Function
Enter diagnostic Signal I/O mode
Exit diagnostics mode
Prints all Signal I/O values. See Table
5.2 for Signal Definitions
Examines or sets I/O signal. See
Table 5.2 for a list of signals. Must
issue D ENTER SIG before using this
command.
Reset analyzer(same as power on)
Resets analyzer and erases RAM.
Erases all DAS data. Keeps setup
variables and calibration
Resets analyzer and erases RAM and
EEPROM. Erases all DAS data.
Resets all setup variables to factory
default. Resets calibration values.