ATIMER
EXITZR
EXITSP
EXITMP
SLPCHG
REPORT PERIOD
This property specifies the period between reports (DAS entries). Enter the number of
days and press ENTR. Then enter the number of hours (the leftmost digits) and minutes
and press ENTR.
NUMBER OF RECORDS
This property specifies how many records you want to store in this data channel. Each
record consists of a time stamp and the sampled measurements from each parameter. The
memory of the M300 has capacity for a total of about 8,000 records (for all data
channels). When the specified number of records for a data channel has been reached,
the channel will "wrap around", with the oldest records being overwritten with the newest
records.
Changing this property requires that any existing data records
be deleted because memory must be re-initialized.
Enter the number of records desired and press ENTR. Only when you press ENTR is
the data file actually deleted. If you press EXIT the property will not be modified and
data file will not be deleted.
RS-232 REPORT
This property controls printing of reports on the RS-232 channel. If set to ON then every
time a new report is stored it's also printed on the RS-232 channel. If this property is set
to OFF, the reports are still stored in the data file, but not printed on the RS-232 channel.
PARAMETERS
This property shows the number of parameters being sampled. Each data channel can
sample up to 10 parameters, each with a separate sampling mode.
This parameter also specifies the mode of data collection. When sampling a reading such
as a concentration at periodic intervals, it is desirable to use the AVG mode to generate an
average of the readings during the interval. But when sampling the concentration when
P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 44
Triggering Events
Description
Automatic timer expired
Exit zero calibration
Exit span calibration
Exit multi-point calibration