Teledyne 300 Instruction Manual page 70

Carbon monoxide analyzer
Hide thumbs Also See for 300:
Table of Contents

Advertisement

"V VARIABLE=VALUE WARNLO WARNHI [DATALO-DATAHI] <CRLF>"
which should reflect the new value. The values in square brackets are not required for all
variables. If needed, the values are included on the command line, separated by spaces.
For example, to change the instrument ID, the host computer would issue a command like
this:
and the CPU should respond with:
"V DDD:HH:MM IIII MACHINE_ID = 1234 (0-9999)<CRLF>"
Table 7.9 lists the variable names which are variable through the RS-232 interface and
their corresponding button sequences.
Var. Name
CO_SPAN
FILT_ADAPT
REPORT_FREQ
MACHINE_ID
BAUD_RATE
CURR_TIME
CURR_DATE
CLOCK_ADJ
FAULT_TIME
P/N 02163G Teledyne API Model 300 CO Analyzer Instruction Manual - Page 64
"V MACHINE_ID=1234<CRLF>"
RS-232 VARIABLE NAMES
Button Sequence
SETUP-IZSC-SPAN
SETUP-MISC-FILT-ADAPT
SETUP-DAS-FREQ
SETUP-COMM-ID
SETUP-COMM-BAUD
SETUP-MISC-CLK-TIME
SETUP-MISC-CLK-DATE
SETUP-MISC-CLK-ADJ
SETUP-MISC-MORE-FLT
TABLE 7.9
Legal Values
0-1000
OFF, ON
1-60
0000-9999
300,1200,2300
00:00-23:59
01/01/00-12/31/99
-60 - +60
0-300

Advertisement

Table of Contents
loading

Table of Contents