Diagnostic Commands And Messages; Table 5-19: Rs-232 C - Teledyne 200A Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API Model 200A NO
Table 5-19: RS-232 Calibration Messages
Message
START ZERO CALIBRATION
1
NOX
=xxxxx PPB
FINISH ZERO CALIBRATION
START SPAN CALIBRATION
1
NOX
=xxxxx PPB
FINISH SPAN CALIBRATION
START MULTI-POINT CALIBRATION
1
NOX
=xxxxx PPB
FINISH MULTI-POINT CALIBRATION
1
Depends on software options installed.
2
Depends on which units are currently selected.

5.5.6 DIAGNOSTIC Commands and Messages

When Diagnostic mode is entered from the RS-232 port, the diagnostic mode issues additional
status messages to indicate which diagnostic test is currently selected. Examples of Diagnostic
mode messages are:
D DDD:HH:MM IIII ENTER DIAGNOSTIC MODE
D DDD:HH:MM IIII EXIT DIAGNOSTIC MODE
Example of turning on the Ozone Generator via the RS-232 port:
D ENTER SIG
D OZONE_GEN=ON
D EXIT
Analyzer Instruction Manual, 02246, Rev. F1
X
2
1
2
NO
=xxxxx PPB
NO2
2
1
2
NO
=xxxxx PPB
NO2
2
1
2
NO
=xxxxx PPB
NO2
Description
Beginning IZS zero calibration
1
2
=xxxxx PPB
Finished IZS zero calibration
Beginning IZS span calibration
1
2
=xxxxx PPB
Finished IZS span calibration
Beginning multi-point calibration
1
2
=xxxxx PPB
Finished multi-point calibration
5-36