Teledyne API Model 200A NO
Table 7-7: Enabling Dynamic Zero/Span
StepNumber
1.
2.
3.
4.
5.
6.
If you try a Dynamic span calibration using the NO
as a calibration gas source, the NO channel concentration will be zero
due to the lack of NO in the NO
will be displayed, CANNOT CALIBRATE. We recommend a calibrated
NO tank for complete calibration. Also see the CAL-ON-NO2 option.
With dynamic calibration turned on, the instrument will re-set the slope and offset values for the
NO and NO
channel each day. This continual re-adjustment of calibration parameters can often
x
mask subtle fault conditions in the analyzer. It is recommended that if Dynamic Cal is enabled,
the TEST functions, and SLOPE and OFFSET values in the M200A are checked frequently to
assure high quality and accurate data from the instrument.
Analyzer Instruction Manual, 02246, Rev. F1
X
Action
Comment
Press
Causes the M200A go to the AutoCal menu.
SETUP-ACAL
Press
Select the sequence you want for dynamic calibration.
PREV-NEXT
Press SET
Select the SET menu.
Press
Scroll through the SET menu to the CALIBRATE attribute.
PREV-NEXT
Press EDIT
Set the CALIBRATE attribute value to ON to enable Dynamic
Span.
Press EXIT
Causes the M200A to return to SAMPLE mode.
NOTE
2
permeation tube
2
permeation tube. An error message
7-8