Table 7-13: Expected Span; Table 7-14: Span Calibration - Teledyne 200A Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API Model 200A NO
Table 7-13: Expected Span Gas Concentration Procedure
Step Number
1.
2.
3.
4.
5.
Sample the generated concentration until the NO and the NO
Span the instrument by the following procedure:
Table 7-14: Span Calibration Procedure
Step Number
1.
2.
3.
4.
5.
The analog voltage output should measure 80% of the voltage range selected. (e.g. 4.00 VDC if
0-5V output is selected.) The readings on the front panel display should be equal to the expected
NO and NO
concentrations entered in the procedure given in Table 7-13 above. See the
x
Troubleshooting Section 9.2.8 if there are problems. Also see the Calibration Quality Check
procedure Section 7.10.
After the zero and the 80% URL points have been set, generate five approximately evenly
spaced calibration points between zero and 80% URL without further adjustment to the
instrument. Allow the instrument to sample these intermediate concentrations for about 10
minutes each and record the instrument NO and NO
Analyzer Instruction Manual, 02246, Rev. F1
X
Action
Comment
Press
This key sequence causes the M200A to prompt for the
CAL-CONC-NOX
expected NO
Enter the NO
under each digit until the expected value is set. This menu can
also be entered from CALS or CALZ.
Press ENTR
ENTR stores the expected NO
Press
Now enter the expected NO span concentration as in step one.
CAL-CONC-NO
Press ENTR
Pressing ENTR stores the NO span value and returns the prompt
to the CONC menu.
Press EXIT
Returns instrument to SAMPLE mode.
Action
Comment
Press CAL
The M200A enters the calibrate mode from sample mode.
Wait 10 min
Wait for readings to stabilize at span values.
Press SPAN
If you change your mind after pressing SPAN, you can still
press EXIT here without spanning the instrument.
Press ENTR
Pressing ENTR actually changes the calculation equations.
Press EXIT
M200A returns to SAMPLE mode.
concentration.
x
span concentration value by pressing the key
x
responses have stabilized.
x
responses.
x
7-23
span value.
x

Advertisement

Table of Contents
loading

Table of Contents