Teledyne 200A Instruction Manual page 156

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API Model 200A NO
Table 9-4: Diagnostic Mode - Signal I/O (Continued)
No.
Signal
10
BROWNOUT_RST
11
CONV_HEATER
12
O3GEN_STATUS
13
AUTO_ZERO_VALVE
14
PREAMP_RANGE_HI
15
ST_RCEL_PRESS
16
ST_ZERO_CAL
17
ST_SPAN_CAL
18
ST_FLOW_ALARM
19
ST_TEMP_ALARM
20
ST_DIAG_MODE
(table continued)
Analyzer Instruction Manual, 02246, Rev. F1
X
Control
Description
YES
Brownout reset works in conjunction with
DSP_BROWNOUT. When DSP_BROWNOUT is
set, the CPU sends a signal to reset the display and
clear the DSP_BROWNOUT.
YES
Shows the status of the Moly heater. This has the
same function as the LED in the Power Supply
Module.
YES
Switches ON/OFF power to the ozone generator.
YES
Switches the AutoZero valve. Use this bit to test the
valve function.
YES
Switches the preamp(01105) hardware range.
Standard ranges are 2000 and 20,000 ppb. Logic high
= 20,000 ppb; logic low = 2000 ppb. M200A will
reset range to correct value based on user set range
value.
YES
Status Bit - Reaction Cell Pressure alarm
Logic High = pressure out of acceptable range
Logic Low = pressure inside acceptable range
YES
Status Bit - Zero Calibration mode
Logic High = M200A in Zero cal mode
Logic Low = Not in Zero cal mode
YES
Status Bit - Span Calibration mode
Logic High = M200A in Span cal mode
Logic Low = Not in Span cal mode
YES
Status Bit - Flow alarm
Logic High = Sample/Ozone flow out of spec
Logic Low = Flows within spec
YES
Status Bit - Temperature alarm
Logic High = Rxcell, Moly, IZS, Box temps out of
spec
Logic Low = Temps within spec
YES
Status Bit - In Diagnostic mode
Logic High = M200A in Diagnostic mode
Logic Low = Not in Diag mode
9-14