1.
In SAMPLE mode, scroll the TEST functions to SAMP FLW. Observe the reading. Then
subtract the observed reading from the desired reading.
2.
Go to the VARS menu by SETUP-MORE-VARS, then pressing ENTR. Scroll to
SFLOW_SET and press EDIT. The first value displayed is the SFLOW_SET:VALUE
reading, ADD the value from step 1 to the reading shown and key in the new reading, then
press ENTR. Check the low and high warning limits which are displayed as the next 2 values
to make sure the new value does not exceed the warning limits. Press exit to return to
SAMPLE mode.
3.
Observe the SAMP FLW value. If necessary repeat step 2 to adjust the reading again to
match the desired flow rate.
Analyzer Instruction Manual, 02246, Rev. F1