Teledyne 200A Instruction Manual page 20

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API Model 200A NO
Step 2 - Calibrate the instrument:
Zero/Span Calibration Procedure
Step Number
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
Analyzer Instruction Manual, 02246, Rev. F1
X
Action
Comment
Input Zero gas
Allow Zero gas to enter the sample port on the rear of the
instrument.
Press CAL
The M200A enters the calibrate mode from sample mode.
Wait 10 min
Wait for reading to stabilize at the zero value. If you wait less
than 10 minutes the final zero value may drift.
Press ZERO
The ZERO button will be displayed.
Press ENTR
Pressing ENTR actually changes the calculation equations and
zeroes the instrument.
Input Span Gas
Switch gas streams to span gas.
Wait 10 min
Wait for reading to stabilize at the span value. If you wait less
than 10 minutes the final span value may drift.
Press SPAN
The SPAN button should be displayed. If it is not, check the
Troubleshooting Section 9.2.8 for instructions on how to
proceed. In certain circumstances at low span gas
concentrations (<100ppb), both the ZERO and SPAN buttons
will appear.
Press ENTR
Pressing ENTR actually changes the calculation equations so
that the concentration displayed is the same as the expected
span concentration you entered above, thus spanning the
instrument.
Press EXIT
Pressing EXIT returns the instrument to SAMPLE mode.