Teledyne 200A Instruction Manual page 106

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API Model 200A NO
If t
is >2 minutes or if P
r
(F
, F
T
O
have to be made, and t
9.
After equations 7-5 and 7-6 are satisfied, calculate the diluent air flow (F
Equation
7-7:
=
F
F
D
T
7.8.4.3 Example GPT Calculation
Following is an example calculation that can be used to determine whether an existing GPT
calibrator will meet the required conditions for a specific calibration. For this example, it is
assumed that only the volume of the reaction chamber, V
standard, [NO]
uses, these flow settings are known and need only to be substituted in Equations 7-5 and 7-6 to
verify the required conditions. Before doing any calculations, the URL and flow demand of the
analyzer being calibrated must be known. Operating parameters are determined from the
operations manual:
Upper range limit = 0.5 ppm, and
Flow demand = 500 cm
Volume of calibrator reaction chamber is determined by physical measurement:
3
V
= 180 cm
RC
The concentration of the NO standard gas to be used is determined by reference against an
NIST-SRM (Section 2.0.7, Q.A. Handbook):
[NO]
= 50.5 ppm
STD
1.
Determine the minimum total flow (F
1:
F
= 500 cm
T
Analyzer Instruction Manual, 02246, Rev. F1
X
is <2.75 ppm-min, changes in flow conditions
R
, F
) or in the reaction chamber volume (V
NO
r
F
F
Equation 7-7
O
N
O
, are known. All flow settings (F
STD
3
/min.
3
/min (110/100) = 550 cm
NOTE
and P
will have to be re-calculated.
R
RC
, F
, F
NO
O
) required at the output manifold using Equation 7-
T
3
/min
7-18
), or both will
RC
D
, and the concentration of the NO
, and F
) will be calculated. In many
T
D
) using

Advertisement

Table of Contents
loading

Table of Contents