Teledyne 200A Instruction Manual page 86

Nitrogen oxides analyzer
Table of Contents

Advertisement

Teledyne API Model 200A NO
Analyzer Instruction Manual, 02246, Rev. F1
X
If an ammonia permeation tube is being used, substitute the molecular weight of NH
(17
3
g/mole) for that of NO
.
2
o
The temperature is set at 50.0
C. Check SETUP-MORE-VARS and scroll to the IZS-SET
o
variable to verify that the temperature is properly set. It should be set to 50
C with over-and-
o
o
under temperature warnings set at 49
C and 51
C. There is a 60 cc/min flow across the
permeation tube at all times to prevent build-up of NO
gas in the tubing.
2
If desired, the output of the permeation tube can be adjusted by adjusting the oven temperature
o
up or down slightly. The adjust increment is 0.1
C to facilitate small adjustments of the setpoint
temperature.
WARNING
Do not leave instrument turned off for more than 8 hours
without removing the permeation tube. Do not ship the
instrument without removing the permeation tube. The
tube continues to emit NO
, even at room temperature
2
and will contaminate the entire instrument.
6-6