Check the TEST functions:
A. Compare the TEST functions to the factory values in Table 2-1. This will often provide
important clues as to the problem.
B. Pay special attention to the NO and NO
C. The slopes are the software equivalent of the span pot on an analog instrument. If the
slopes are not 1.0 ±0.3, the gain has changed. Do the following: Check the AutoZero
reading in the TEST functions. Compare it to the value in the factory checkout Table 2-1.
If the reading is significantly greater than the factory test value, the reaction cell could be
contaminated or there could be a light leak in the cell. Verify this fault by turning the
ozone generator off and see if the reading drops more than 25 mV.
D. Check for a change in cell pressure(vacuum) - compare to value in Table 2-1. - possible
causes:
1)
2)
3)
4)
4.
Check for pneumatic leaks - perform the leak check procedure in Section 8.9 If slopes are
different from each other by > .1, this usually indicates a leak in the switching (NO/NO
valve or improper calibration.
5.
Check for light leaks - Turn off the ozone generator, then wait 7 minutes. If the reading
drops significantly, the reaction cell is contaminated. If not a light leak is indicated.
6.
Incorrect span gas concentration - this could come either from the calibrator or entering
the expected span gas concentration in the M200A incorrectly, see Table 7-3.
7.
If the instrument does not respond to span gas, check Section 9.2.3.
The above should get you started in diagnosing and repairing the most common faults. If these
reasons have been eliminated, the next thing to do is a Factory Calibration covered in Section
9.1.6 or check Section 9.2 for other fault diagnosis. If difficulties persist, contact our service
department. The 800 telephone number is on the cover page of this manual.