Zero Calibration; Span Calibration; Figure 8-1: Gpt Calibration System - Teledyne 200E Instruction Manual

Chemiluminescence nitrogen oxides analyzer
Table of Contents

Advertisement

Model 200E Instruction Manual
EPA Protocol Calibration

Figure 8-1: GPT Calibration System

Converter efficiency should be set prior to calibration since its value is used in the
computation of the NO
and NO
concentration outputs.
X
2
The analyzer should be calibrated on the same range used for monitoring.
If auto ranging or independent range options are selected the highest of the ranges will
result in the most accurate calibration, and should be used.
Make sure the GPT calibration system can supply the range of concentrations at a sufficient
flow over the whole range of concentrations that will be encountered during calibration.

8.3.1. Zero Calibration

The zero point calibration procedure is identical to that described in Sections 7.2, 7.4 or 7.5
of this manual.

8.3.2. Span Calibration

Adjust the NO concentration to approximately 80% of the URL of the NO channel. The
expected NO and NO
span concentrations can be determined by measuring the cylinder
X
and diluent flows and computing the resulting concentrations. If there is any NO
impurity
2
in the NO standard gas it should be taken into account when the NO
concentration is
X
entered during the NO/NO
channel calibration. This is done by ADDING the impurity
X
044100102 Rev A
133