Confirm that the image is
used for radiotherapy
localization purpose before
implementing the procedures
on the right using the read-
only cassette.
p.36 Erasing the
REGIUS Plate
p.26 Name of Parts
p.27 Cassette Structure
Loading of the REGIUS
Plate in the read-only
cassette should always be
carried out in dark room.
REGIUS MODEL 190 OPERATION MANUAL Ver. 1.00 2004/10
Usage of the Exposure-Only Cassette
(images for the radiotherapy localization)
Procedures ( "Loading the plate in the read-only cassette" -> "Exposure" -> "Loading the plate
in the system cassette" ) to record the exposed image data are described below.
Open/Close
Lock
Back Panel of
Cassette
Front Panel of
Cassette
Back Panel
of Cassette
Read-only cassette
[Plate Release]
button
REGIUS Plate
"UPPER" label
1
Erase the images on the REGIUS Plate of
the read-only cassette before exposure.
2
Disengage the open/close lock of the read-
only cassette, and separate the front plate
from the back plate on which the REGIUS
Plate is attached.
3
Align the register mark of the back plate to
that of the read-only cassette, and set the
back plate in the read-only cassette.
4
Lift the back plate while pressing the [Plate
Release] button to release the REGIUS
Plate, and set the REGIUS Plate in the
exposure-only cassette.
5
Close the exposure-only cassette's lid, and
make sure the lock is engaged.
6
Implement an exposure so that the frame
on which "UPPER" label is affixed comes
to the top of the image.
7
After the exposure is completed, open the
lid of the exposure-only cassette in the
dark room. Set the back plate of the read-
only cassette with its register mark aligned
to the register mark of the exposure-only
cassette. Pull out the REGIUS Plate by
making the REGIUS Plate stick together
with the back plate.
8
Couple the front and back plates of the
read-only cassette, and make sure the lock
is engaged.
9
Insert the read-only cassette into the
REGIUS MODEL 190 to read the image
data.