Emerson Rosemount 3308 Series Reference Manual page 113

Wireless guided wave radar
Table of Contents

Advertisement

Reference Manual
00809-0100-4308
Table 6-4: Maintenance Alerts (M:)
Message
Non-PV Out of Limits
Table 6-5: Advisory Alerts (A:)
Message
Database Memory Warning
Non-Critical User Data Warning
Volume Range Warning
Verification Mode Active
Button Stuck
Reference Manual
(continued)
Description
One of the non-primary
measurements is outside the
associated sensor limits and may be
unreliable.
Description
The device has failed to write to the
database memory at some time in the
past. Any data written during this time
may have been lost.
A user written parameter does not
match expected value.
The level measurement is outside the
configured volume range.
The device is in verification mode and
is not reporting actual information.
The button on the Electronics Board is
detected as stuck in the active
position.
Service and troubleshooting
February 2020
Recommended actions
1. Bring the system to a safe state.
2. Verify that all non-primary
measurements are within
specified limits.
3. Restart the device.
4. If the condition persists, replace
the device
Recommended actions
1. If logging dynamic data is not
needed, this advisory alert can be
safely ignored.
2. Restart the device.
3. Reconfirm all configuration items
in the device.
4. Restore default settings and
reconfigure the device.
5. If the condition persists, replace
the device.
1. Restart the device.
2. Reconfirm all configuration items
in the device.
3. Restore default settings and
reconfigure the device.
4. If the condition persists, replace
the device.
Check volume configuration.
If this behavior is not desired, stop
verification mode.
1. Check the buttons for
obstructions.
2. If the condition persists, restart
the device.
3. If the condition persists, replace
the device.
113