Service and troubleshooting
February 2020
Table 6-3: Configuration Error Details (D:)
Message
Lower Range Value or Upper Range
Value is out of limits
Configured Measurement Mode not
Supported
Volume Configuration Error
Unsupported Probe and Nozzle
Configuration
Parameter Out of Limits
Low Low Level Alert configuration is
invalid
Low Level Alert configuration is invalid Low Level Alerts will not be raised as
High Level Alert configuration is invalid High Level Alerts will not be raised as
High High Level Alert configuration is
invalid
110
Description
The lower/upper range value is outside
the lower/upper sensor limits. This is
outside the range where the sensor
works properly, hence the
measurement may be unreliable.
The configured Measurement Mode
does not work since support has not
been purchased.
The volume cannot be calculated
correctly with the current
configuration.
Rigid Single Lead 0.5 in / 13 mm probe
is configured in combination with 1.5
inch Nozzle. This combination is
invalid and results in unreliable
measurement.
One or more of the configuration
values, in a set that belongs together,
has been changed and reduced the
measurement range of the device.
Low Low Level Alerts will not be raised
as expected because of invalid
configuration.
expected because of invalid
configuration.
expected because of invalid
configuration.
High High Level Alerts will not be
raised as expected because of invalid
configuration.
Reference Manual
00809-0100-4308
Recommended actions
Check the lower and upper range
values in relation to the sensor
limits.
1. Upgrade the device.
2. Change the Measurement Mode
to match, refer to
Measurement
mode.
1. Check that the level-volume
values in the strapping table are
entered in increasing order, refer
to Volume.
2. Check that the number of
strapping points to use is correct.
3. Check size measures in the
Volume Setup.
Install a nozzle with a supported
inner diameter and re-configure
transmitter.
1. Check the configured values for
Probe Length, Vapor Dielectric
Constant, and Upper Product
Dielectric Constant.
2. Restore default settings and
reconfigure the device.
Check entered limit and deadband
values in comparison with entered
tank and probe parameters such as
probe length, tank height etc.
Check entered limit and deadband
values in comparison with entered
tank and probe parameters such as
probe length, tank height etc.
Check entered limit and deadband
values in comparison with entered
tank and probe parameters such as
probe length, tank height etc.
Check entered limit and deadband
values in comparison with entered
tank and probe parameters such as
probe length, tank height etc.