Esd Precautions; Tools Required - Agilent Technologies 54830 Series Service Manual

Table of Contents

Advertisement

Replacing Assemblies
Use the procedures in this chapter when removing and replacing assemblies and parts
in the Agilent Technologies 54830 Series oscilloscope.
In general, the procedures that follow are placed in the order to be used to remove a
particular assembly. The procedures listed first are for assemblies that must be removed
first.

ESD Precautions

When using any of the procedures in this chapter you must use proper ESD precautions.
As a minimum you must place the oscilloscope on a properly grounded ESD mat and
wear a properly grounded ESD wrist strap.
CA UT IO N
AVOID DAMAGE TO THE OSCILLOSCOPE!
Failure to implement proper antistatic measures may result in damage to the oscilloscope.

Tools Required

The following tools are required for these procedures.
• Torx drivers: T6, T8, T10, T15, T20
• Socket wrench: 5/8 inch
• Medium size (3/16-in) flat-blade screwdriver
• Nut Drivers: 3/16-in, 9/32-in, 5/16-in, 5/8-in
• Torque driver, 0.34 Nm (3 in-lbs), 5 mm or 3/16-in hex drive
• Torque driver, 0.34 Nm (3 in-lbs), Torx T6 drive
CA UT IO N
REMOVE POWER BEFORE REMOVING OR REPLACING ASSEMBLIES!
Do not remove or replace any circuit board assemblies in this oscilloscope while power is applied.
The assemblies contain components which may be damaged if the assembly is removed or
replaced while power is connected to the oscilloscope.
WAR N IN G
SHOCK HAZARD!
To avoid electrical shock, adhere closely to the following procedures. Also, after disconnecting
the power cable, wait at least three minutes for the capacitors on the power supply to discharge
before servicing this oscilloscope. Hazardous voltages exist on the inverter for the display monitor
while power is applied.
6–2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

54833a54830b54830d54831b54831d54832b ... Show all

Table of Contents