Chapter 3: Testing Performance
To test time measurement accuracy
Select horizontal from the setup menu. Set the scale to 5 ns/div, position at –5.0 ns, and
8
reference to the left.
See Figure 3-8.
Figure 3-8
Horizontal Setup for Equivalent Time Procedure
Adjust the signal generator output voltage to obtain a waveform with a rise time of
9
approximately 1.4-2.0 ns.
You can measure the rise time by selecting the rise time command from the Time submenu of
the Measure menu. See Figure 3-9.
Figure 3-9
Use the rise time
measurement to verify
correct waveform shape
for the time
measurement accuracy
check
Waveform for Time Measurement Accuracy Check
Select Acquisition from the Setup menu.
10
3–16