To disconnect the probe, push the small latch on top of the probe connector to the left,
3
then pull the connector body away from the front panel of the oscilloscope without
twisting it.
Figure 2-7
Disconnecting the Oscilloscope Probe
CA UT IO N
Do not attempt to twist the snap-on probes on or off the oscilloscope's BNC connector. Twisting
the probe connector body will damage it.
CA UT IO N
For the 54830 series oscilloscopes do not exceed the maximum input voltage rating. The
!
maximum input voltage for 50 Ω inputs is 5 Vrms, CAT I. Maximum voltage at 1 MΩ impedance
is ±100 V (dc + ac) [ac < 10 kHz], CAT I.
Chapter 2: Preparing for Use
To connect oscilloscope probes
2–11