Quality Assurance
Setting scan and recon param-
eters
syngo CT 2007E
02/2008
You must set special scan and recon parameter to perform the
low contrast test adequately.
F . 5
Scan parameter
mAs
kV
Scan time
Slice
No. of scans
Recon parameter
Kernel
FoV
Enter or select the required parameter in the subtask cards
Routine and Recon.
Low Contrast Test (Optional)
F . 5
130
F . 5
130
F . 5
1.0 s
F . 5
10.0 mm
F . 5
1
F . 5
F . 5
B20s
F . 5
150 mm
F . 5
Emotion Operator Manual
F . 5
F . 5
F . 5
F . 5
F . 5
F . 5
F . 5
F . 5
F . 5
F . 5
F . 5
F . 5
0 . 0
F.5-11
0 . 0
0 . 0