Siemens SOMATOM Emotion syngo CT 2007E Operator's Manual page 337

Table of Contents

Advertisement

Operation
Checking the table top
Checking the laser light marker
syngo CT 2007E
02/2008
You must check the mobility and cleanness of the table top.
D . 2
Make sure that you can manually retract the table top from
the gantry. (→ Page C.2-9 Retracting the patient table
manually)
Make sure that the scan field is free of residual contrast
medium, blood or other contaminations.
D . 2
Switch on the laser light marker and check the projections
of the light beams on a white sheet of paper.
The laser beam must project a line.
D . 2
The laser light marker does not work.
f B
D . 2
Stop scanning to rule out any danger to patients.
Call Siemens Customer Service.
Function Tests
Emotion Operator Manual
D . 2
D . 2
D . 2
D . 2
0 . 0
D.2-3
0 . 0
0 . 0