Quality Assurance
Starting the test
syngo CT 2007E
02/2008
F . 4
In the Quality Constancy dialog box, select the individual
tests CTDI Head (opt.) and CTDI Body (opt.).
(→ Page F.3-2 Performing the constancy test)
F . 4
Confirm selection of the constancy test with Go.
F . 4
You are prompted to enter the name of the tester and the serial
numbers of the phantoms.
Enter your name and the data required.
Start the CTDI test with Go.
F . 4
If you are performing the CDTI measurements as part of the
i
F
. 4
monthly constancy test, replace the set of phantoms with the
CTDI head phantom used for the constancy test now.
Confirm the correct phantom position with Go.
F . 4
You are prompted to press Start.
CTDI Test
Emotion Operator Manual
F . 4
F . 4
F . 4
F . 4
0 . 0
F.4-5
0 . 0
0 . 0