Quality Assurance
Triggering radiation
syngo CT 2007E
02/2008
In the Quality Daily dialog box you are prompted to initiate
F . 2
radiation.
Press the Start key on the control box.
The position of the phantom set is checked and displayed. Then,
the first measurement is started automatically.
F . 2
For incorrect positioning of the phantom set, a message is dis-
i
F
. 2
played. Correct the position, if necessary.
For each tube voltage (kV step), two measurements are per-
formed.
Daily Quality Measurements
Emotion Operator Manual
F . 2
F . 2
F . 2
F . 2
0 . 0
F.2-5
0 . 0
0 . 0