Daily Quality Measurements
Activating daily quality test
Starting measurement
F.2-4
syngo CT 2007E
02/2008
Call up Setup > Quality.
F . 2
The Quality Daily dialog box is displayed.
F . 2
Click on the OK button.
The tilt of the gantry is automatically set to 0°, if necessary.
F . 2
The "Quality Assurance Patient" is automatically selected.
The patient table is positioned such that the water phantom is
in the scan plane.
Quality Assurance
Emotion Operator Manual
F . 2
F . 2
F . 2
F . 2
F . 2
0 . 0
0 . 0